0%

MODEL 570 PRESSURE TRANSMITTER

4-20 mA, 0-100 psi thru 100K psi, Hazardous Approvals

570 provides a rugged, sealed solution for hazardous rated and harsh environments.

Features

 Many Options available

 All Stainless Steel construction

 IP-68 Rated

 ShuntCal

 High Pressure Applications

 RapidShip – 2 Day Turnaround

All welded construction and stainless steel surfaces makes the 570 ideal for corrosive environments.

The transmitters O-ring free design ensures reliability with corrosive media.

Viatran’s 570 FM approved labelled units are designed to meet NEC/API Dual Seal requirements.

For the most corrosive environments, Hastelloy, Inconel, and 316 SS wetted parts are available.

Full Scale Pressure Range:  0-100 thru 0-100K PSIS (0-6.89 thru 0-6894 bar)

Accuracy (RSS):

(BFSL Non-Linearity, Hysteresis, Non-Repeatability)

  • 0-300 PSI thru 0-15K PSI Ranges±0.20% FSO ( ±0.15% FSO improved with “DN” option)
  • < 0-300 PSI or > 0-15K PSI Ranges±0.25% FSO ( ±0.15% FSO improved with “DN” option)

Linearity (Best Fit Straight Line)

  • 0-300 PSI thru 0-15K PSI Ranges±0.10% FSO (BFSL), Typical
  • < 0-300 PSI or > 0-15K PSI Ranges±0.15% FSO (BFSL), Typical

Hysteresis:  < ± 0.05% FSO, Typical

Repeatability:  < ± 0.05% FSO, Typical

Full Scale Output (FSO):  16 mA

Zero and Span Balance:  ±1% FSO

Long Term Stability:  ≤±0.25% FSO per 6 months

Response Time:  <2 mSec to reach 90% FSO

Temperature Effect on Zero:  ≤±1% FSO per 100°F (37°C)

Temperature Effect on Span:  ≤±1% FSO per 100°F (37°C)

Compensated Temperature:  70°F to 170°F (21°C to 76°C)

Operating Temperature:  -40°F to 190°F (-40°C to 87°C)

Storage Temperature Limit:  -65°F to 250°F (-53°C to 121°C)

Supply Voltage:  12 to 30 Vdc

Power Supply Regulation Effect:  ≤±0.02% FSO per Volt

Output Signal:  4 – 20 mA

Bridge Impedance:  5000 Ohms nominal

Circuit Protection:  Varistor protected across the input leads for surges above 40 V and currents to 250 Amps peak with a pulse width of 20 μSecs. Reverse polarity protected.

Insulation Resistance:  >200 MegOhms to case ground

Electrical Connection:  1/2″ NPT (M), 18 AWG wire, 72″

Wire Color Codes:

Red          +Power/Signal

Black        -Power/Signal

Green         Case ground

White         No Connection

Pressure Connection:

  • 0-50 thru 0-15K PSI  (1/4″ – 18 NPT Female (0-3.3 thru 0-1034 bar))
  • 0-20K thru 0-50K PSI  (1/4″ F250-C High Pressure Tube (0-1378 thru 0-3447 bar))
  • 0-60K thru 0-100K PSI  (5/16″ F312C High Pressure Tube (0-4136) thru 0-6894 bar))

Proof Pressure:

  • 0-100 thru 0-15K PSI  (1.5 times FSPR or 20K PSI (1378 bar), (0-6.8 thru 0-1034 bar) whichever is less)
  • 0-20K thru 0-100K PSI  (1.2 times FSP (0-1378 thru 0-6894 bar))

Burst Pressure:

  • 0-100 thru 0-15K PSI  (≥ 2.3 to 5 times FSPR dependent on (0-6.8 thru 0-10342 bar) sensor)
  • 0-20K thru 0-100K PSI  (1.5 times FSPR or 125K PSI (8618 bar) (0-1378 thru 0-6894 bar) whichever is less)

Mounting:  Supported by process piping or optional mounting bracket

Enclosure:

  • 0-100 thru 0-15K  (PSI15-5 PH and 316 stainless steel (0-6.8 thru 0-1034 bar))
  • 0-20K thru 0-100K  (PSI316 SST and PH 13-8 Mo SST (0-1378 thru 0-6894 bar))

Wetted Parts:

  • 0-100 thru 0-15K  (PSI15-5 PH stainless steel (0-6.8 thru 0-1034 bar))
  • 0-20K thru 0-100K PSI  (PH 13-8 Mo SST (0-1378 thru 0-6904 bar))

Shock Limitation:  100 G’s

Weight:  24 oz. (0.680 kilograms)

Identification:  Laser etched onto body

Enclosure:  Classification NEMA 4X

DF:  Bleed port (10K PSI and below)

DG:  Improved temperature compensation (<± 0.5% FSO/100°F for zero/span)

DH:  Special range

DN:  Improved Accuracy (±0.15% FSO)

DQ:  Cleaning for oxygen service

EA:  Special calibration run

JJ:  Direct mount indicator with explosion proof housing

JS:  Direct mount indicator

KN:  IECEx flameproof label

ME:  CSA explosion proof label

MR:  Modified electrical damping

NG:  ATEX flameproof label

NH:  Customer specified identification

NJ:  CE label

NK:  ATEX intrinsic safety label

NX:  CSA intrinsic safety label

NY:  FM explosion proof label

NZ:  FM nonincendive label

PW:  1/8 DIN digital indicator (consult factory)

QA:  316 stainless steel wetted parts

QB:  Inconel (X-750) wetted parts

QC:  Hastelloy (C-276) wetted parts

TF:  FM intrinsic safety label

Y ( ):  Alternate pressure ports

Note:

Application of some available options may affect standard performance.