4-20 mA, 5″wc to 100 psi
Viatran’s 344 sensor is a low range pressure transmitter designed for the test market. The 344 features highly accurate pressure measurement in ranges from 5 inches water column to 100 PSI. The unique variable capacitance technology allows for extreme overpressure protection up to 500 PSI.
Features
High Accuracy
Many Options available
Low Pressure Applications
The Model 344 offers a 4 to 20 mA signal.
Using our variable capacitance technology, the Model 344 transmitter provides extreme overpressure protection up to 500 PSI.
This sensor design provides standard accuracy of <±0.15% FSO.
The unit also offers standard temperature performance to ±2%/100° F with an improved option to±1%/100° F.
All wetted surfaces of the X44 Series are constructed with 316 stainless steel, making them ideal for wet/wet applications.In environmental test chambers, these surfaces withstand contaminants such as water vapor, oil, exhaust or mists.
Atmospheric reference is specially constructed to withstand the ingress of moisture, ensuring accurate pressure measurement.
Model 344 features an internal calibration circuit for easy field set-up, a quick disconnect electrical connection and external zero and span controls.
These features allow you to calibrate the transducer without an external pressure source and eliminate the need to send the unit to a metrology lab for calibration.
Full Scale Pressure Ranges:
- 0-5, 10, 20, 50, 100, 300, 750″ W.C.
- 0-15, 50, 100 PSIG
Nonlinearity (Best Fit Straight Line): ≤±0.15% FSO
Hysteresis & Repeatability: ≤±0.10% FSO
Full-Scale Output (FSO): 16 mA at 70°F (21°C)
Long Term Stability: ≤±0.2% FSO per 6 months
Zero Shift After 500 PSI Overload: ≤± 0.1% FSO
Compensated Temperature Range: 70°F to 170°F (21°C to 76°C)
Ambient Operating Temperature Range: 0° F to +250° F
Storage Temperature Limits: -40°F to 250°F (-40°C to 121°C)
Temperature Effect on Zero: <± 2.0% FSO per 100° F
Temperature Effect on Span: <± 2.0% FSO per 100° F
Supply Voltage: 10 to 42 Vdc
Power Supply Regulation: ≤±0.1% FSO change over the supply voltage range
Output Signal: 4 to 20 mA
Load Resistance: 1600 Ohms with input voltage change from 10 to 42 Vdc
Zero Adjustment: ± 5 % FSO min
Span Adjustment: ± 10% FSO min
Calibration Signal: 80% of FSPR, by shorting pins C & D
Calibration Signal Accuracy: ≤±0.5% FSO
Circuit Protection: Reverse polarity protected
Insulation Resistance: <1nS
Response Time: From 0.2 Sec for 5″ W.C. range down to 0.008 Sec for 100 PSI range. (Faster response available with NF option.)
Electrical Connection: Bendix PT02E-10-6P, mates PT06E-10-6S (SR)
Pin Outs:
Pin A + Power / Signal
Pin B – Power / Signal
Pin C Calibrate
Pin D Calibrate
Pin E No Connection
Pin F No Connection
Wetted Parts: 316 Stainless steel
Housing: 304 and 316 Stainless Steel with an Aluminum alloy, black zinc-cobalt plated electrical connector
Weight: 3.6 lbs.
Identification: Laser etched onto body
Pressure Connection: 1/4″ – 18 NPT female
Proof Pressure: 500 PSI
Burst Pressure: 1000 PSI
Position Sensitivity: ≤ ± 1.0” W.C. in any orientation
Shock & Vibration: ≤ ± 1.0% FSO per G for 50 Hz to 200 Hz along any axis
Fill Fluid: Dow DC200 Silicone oil
Mounting: May be supported by process piping or by optional mounting bracket
ELECTRICAL CONNECTION:
BA: (K)PTIH-12-8P
BF: (K)PTIH-10-6P
BG: DIN 43650
BI: CF3102E-14S-2P
BK: WK4-32S
BL: WK6-32S
BM: WKA7-32S
BN: (K)PTIH-8-4P
BQ: (K)PT02H-10-6P
BR: CF3102E-14S-6P
BS: CF3102E-14S-5P
ZS: (K)PTO2E-10-6P
ZV: C31-10-6P-20-OP1 (SS)
PERFORMANCE OPTIONS:
DG: Improved temperature compensation to ±1.0% FSO per 100°F
DH: Special ranging
DK: Special shunt calibration setting
DQ: Cleaning for oxygen service
EA: Special calibration run
NF: Fast response time
NH: Customer specified identification
PW: 1/8 DIN digital indicator
ALTERNATE PRESSURE PORTS:
YA: MS 33649-04
YH: 1/8 NPT(M)
YI: 1/8 NPT(F)
YK: 1/4 NPT(M)
YN: 3/8 NPT(F)
Note:
Application of some available options may affect standard performance.