0-5 Vdc output, 0-3 to 20K psi
247 offers a solid, reliable pressure transducer with excellent noise immunity, repeatability and stability.
Features
Model 247’s are designed to provide a voltage output, standard setup is 0-5 Vdc out, but the unit can be configured to provide outputs from -10 to +10 Vdc out and anything in between.
The 247 series can be built to many custom configurations, and with the quality and reliability you expect from Viatran, they are built to last for a very long time.
Full Scale Pressure Ranges:
- 0-15, 30, 50, 75, 100, 200, 300, 500 PSIA
- 0-3, 5, 10, 15, 30, 50, 75, 100, 200, 300, 400, 500, 750, 1,000, 1,500 PSIG
- 2K, 3K, 4K, 5K, 7.5, 10K, 15K, 20K PSIS
- 0-5, 10 & 15 PSIV*
Improved Accuracy (Non Linearity): ±0.35%** (** Range Dependent)
Non-linearity (Best Fit Straight Line): ≤±0.50% FSO
Hysteresis: <0.25% FSO
Full Scale Output (FSO): 5 Vdc ±2% at 70°F (21°C)
Resolution: Infinite
Long Term Stability: ≤±0.5% FSO per 6 months
Compensated Temperature Range: 70°F to 170°F (21°C to 76°C)
Maximum Operating Temperature: 200°F (93°C)
Minimum Operating Temperature: -20°F (-28°C)
Storage Temperature Limits: -40°F to 250°F (-40°C to 121°C)
Thermal Effect on Zero: ≤±2% FSO per 100°F Typical
Thermal Effect on Span: ≤±2% FSO per 100°F Typical
Zero Adjust: ±5% of FSO (nominal)
* For PSIV and Compound Range units, thermal effect on zero is ≤±3.0% per 100°F
Supply Voltage: 8 to 30 Vdc
Power Supply Regulation: ≤±0.05% FSO per Volt
Output Signal:
- 0-5 Vdc
- w/DM option ~ 0 to 10 Vdc (requires 12 Vdc supply min)
Load Impedance: ≥50K Ohm resistive
Circuit Protection: Reverse polarity protected. CE marked
RFI / EMI: CE EMC compliant as per IEC EN 61326-1 and EN 61326-2-3 Annex BB
Voltage Spike: Withstand 1000 volt spike per EN 61000-4-5
Insulation Resistance: <5 nS conductance
Input current: ≤4 mA
Response Time: <2 mSec for 10-90% change in output
PIN OUTS:
Pin 1 +Power
Pin 2 +Signal
Pin 3 Common
Ground Spade Case ground
Pressure Connection: 1/4” NPT female
Proof Pressure:
- ≤200 PSI 3x
- ≥ 300 PSI 1.5 times FSPR or 25K PSI (1723 bar), whichever is less
Burst Pressure: 5 times FSPR or 30K PSI (2068 bar), whichever is less
Pressure Cavity Volume: 0.06 cubic inches, nominal
Wetted Parts:
- 0-3 thru 200 PSI (0-1.0 thru 13.7 bar) – 316L stainless steel
- 0 – 300 PSI (20.6 bar) and above – 15-5PH stainless steel
Weight: 8 oz nominal silicon sensor, 10.5 oz nominal (strain gage)
Identification: Synthetic label
Mounting: May be supported by process piping
DH: Special Range
DM: Modified Full Scale Output (>5 Vdc)
DN: Improved Linearity 0.35% FSO (BFSL) for 0-3 thru 0-20,000 PSI
DX: Modified Full Scale Output (≤5 Vdc)
NH: Customer Specified Marking
Note:
Vented gage units should only be installed in environments that allow the breather to remain dry and free from contamination. Application of some available options may affect standard performance.